Anritsu MG369XB Portable Generator User Manual


 
Chapter 5
Operation Verification
5-1 Introduction This chapter contains three operation verification tests that can be
used to verify Series MG369XB Synthesized signal generator opera
-
tion.
Setup instructions and performance procedures are included for each
test. The results can be compared with the specified limits that are
shown on the test record forms that are provided for each test.
5-2 Test Equipment Table 5-1 lists the recommended test equipment for performing the op
-
eration verification tests in this chapter.
MG369XB OM 5-3
Instrument Critical Specification
Recommended
Manufacturer/Model
Frequency
Counter
or
Range: 0.01 to 40 GHz
Input Z: 50W
Resolution: 1Hz
Other: External Time Base
Input
Anritsu Model MF2414B
Frequency
Counter,
with
Cable Kit
and
External Mixer
Range: 0.01 to 65 GHz
Input Z: 50W
Resolution: 1Hz
Other: External Time Base
Input
EIP Microwave, Inc. Models 538B,
548B, or 578B,
with
Cable Kit: Option 590 and
External Mixer:
Option 91 (26.5 to 40 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)
Power Meter,
with
Power
Sensor
Range: –30 to +20 dBm
(1mW to 100 mW)
Anritsu Models ML2437A or
ML2438A,
with
Power Sensor:
MA2474A (0.01 to 40 GHz)
MA2575A (0.01 to 50 GHz)
Oscilloscope Bandwidth: DC to 150 MHz
Vertical Sensitivity: 2 mV/
division
Horizontal Sensitivity: 50 ns/
division
Tektronix, Inc. Model TAS485
Table 5-1. Recommended Test Equipment