Tektronix AWG610 Portable Generator User Manual


 
Operating Basics
2-38
AWG610 Arbitrary Waveform Generator User Manual
Table 2-14: Setup output parameter operations (cont.)
Bottom
button
Description
Side
button
Horizontal Clock
Clock Ref
Marker 1
Delay
Marker 2
Delay
Sets the clock sample rate from 50 kS/s to 2.6 GS/s.
Sets the reference clock source to either Internal or External. A
valid external clock signal is 10 MHz ±0.1 MHz with a voltage
level of 0.2 to 3.0 V
pĆp
.
Sets the Marker 1 signal delay value from0sto1.5ns.
Sets the Marker 2 signal delay value from0sto1.5ns.
Run Mode Continuous
Triggered
Gated
Enhanced
Displays the Run Mode side menu for setting the instrument run
mode. Refer to the Run Mode Menu section on page 3-37 for an
explanation of the different run modes.
Trigger Source
Slope
(or Polarity)
Level
Impedance
Interval
Sets trigger source to Internal or External. If External selected, all
other side menu items are not selectable except Interval.
Sets the trigger slope or gate polarity to Positive or Negative.
Sets the trigger signal level. The trigger level range is ±5.0Vin
0.1 V increments.
Sets the external trigger input line impedance to either 50 W or
1 kW.
Sets trigger interval from 1.0 ms to 10.0 s.
Save/ReĆ
store
Save Setup
Resore SetĆ
up
Save the setup parameters set by SETUP window and FG mode
window as a setup file.
Restore a setup file.
To output a loaded waveform, push the CH 1 OUT and/or CH 1 OUT front-
panel button(s), then the RUN front-panel button. The LEDs near each button
light up to indicate they are enabled. The instrument outputs the waveform
depending on the Run mode. You can turn either or both channel outputs on or
off while the instrument is running by pushing the CH 1 OUT or CH 1
OUT
buttons. To stop the waveform output, push the RUN button so that the LED
turns off.
Outputting a Waveform