Tektronix AWG610 Portable Generator User Manual


 
The UTILITY Window
3-166
AWG610 Arbitrary Waveform Generator User Manual
Do the following steps to execute the calibration:
1. Push the RUN button to turn the output off if a waveform is being output.
The RUN LED is off.
2. Push UTILITY (front-panel)!Diag (bottom)!Execute Calibration (side).
The internal calibration routine runs immediately and requires up to
15 seconds to complete.
The status message box appears when calibration has been terminated. See
Figure 3–56.
Figure 3-56: Status message box
Pass is displayed in the message box if the calibration successfully terminates.
Fail is displayed if calibration encounters a problem.
Push the OK side button or CLEAR MENU button to erase the status message
box and return to the screen shown in Figure 3–55.
At power on, a limited set of hardware tests for all the test categories are
performed and the results are displayed on the screen. The instrument displays
the following message when an error is detected:
Press any key to go to the SETUP menu screen.
An error in the digital to analog converter (DAC) may be reported if you do not
execute the system calibration at power-on.
See Table 3–45 for the test categories and error codes.
PowerĆon Diagnostics