Tektronix AWG610 Portable Generator User Manual


 
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-21
Waveform
output
CH1
Trigger
signal
CH2
Figure B-7: Relationship between trigger signal and waveform output
Equipment
required
Two 50ĂW BNCĂcoaxial cables, aĂ50ĂW SMAĂcoaxial cable,
ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter,ĂaĂBNCĆTĂ(male to 2 females) adapter, a
function generator, and an oscilloscopeĂ(TDS700).
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
1. Set the oscilloscope controls as follows:
Vertical......................... CH1andCH2
CH1 and CH2 coupling . ........ DC
CH1scale .................. 0.5V/div
CH2scale .................. 2V/div
CH1 input impedance . ......... 50W
CH2 input impedance . ......... 1MW
Horizontal
Sweep . .................... 20ms/div
Trigger
Source . .................... CH1
Coupling .................... AC
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto
2. Set the function generator (AFG310) controls as follows:
Check Gated Mode