Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-37
3. Push UTILITY (front-panel)!System (bottom)!Factory Reset
(side)!OK (side).
4. Load the SINE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
5. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
6. Do the following substeps using the n MKR function of the 497P to check
the harmonics and noise level:
a. Verify that the harmonics level of the spectrum displayed on the
spectrum analyzer from 0 Hz to 800 MHz is -40 dBc or less.
b. Verify that the noise level of the spectrum displayed on the spectrum
analyzer from 0 Hz to 800 MHz is -50 dBc or less.
7. Push the CH1 output button to turn off the CH1 LED.
8. Disconnect the spectrum analyzer.