Appendix B: Performance Verification
B-38
AWG610 Arbitrary Waveform Generator User Manual
Internal Trigger Tests
These procedures check internal trigger function of the AWG610 Arbitrary
Waveform Generator.
NOTE. When you output signal from the CH1 or CH1
OUTPUT, check that the
other OUTPUT ( CH1
or CH1 ) LED is off.
If the other OUTPUT LED is on, push the CH1
or CH1 OUT button to turn off
the output.
Equipment
required
A50W SMAĂcoaxial cable,ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter and an
oscilloscopeĂ(TDS700).
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use the coaxial cable and SMA(Fe)-BNC(Fe) adapter to connect the
AWG610 Arbitrary Waveform Generator CH1 output connector to the
oscilloscope CH1 input connector (see Figure B–14).
OscilloscopeĂ(TDS700)
50 W SMAĂcoaxial cable
AWG610 Arbitrary Waveform Generator
SMA(Fe)ĆBNC(Ma)
adapter
Figure B-14: Internal trigger initial test hookup
2. Set the oscilloscope controls as follows:
Vertical . ........................ CH1
CH1 coupling . ............... DC
CH1 scale . ................. 0.5V/div
CH1 input impedance . ......... 50W
Horizontal