Appendix B: Performance Verification
B-48
AWG610 Arbitrary Waveform Generator User Manual
8. Check the EVENT IN connector pin 1 input by following the substeps
below:
a. Close SW2 of the ground closure to generate an event signal on the
EVENT IN connector pin 1.
b. Verify that the oscilloscope displays the waveform as shown in Figure
B–23.
Figure B-23: Waveform output when SW2 is closed
c. Open the SW2 of the ground closure to degenerate the event signal.
d. Verify that the oscilloscope displays the waveform shown in Figure
B–21.
9. Check the EVENT IN connector pin 2 input by following the steps below:
a. Close SW3 of the ground closure to generate an event signal on the
EVENT IN connector pin 2.
b. Verify that the oscilloscope displays the waveform shown in
Figure B–24.