Tektronix AWG610 Portable Generator User Manual


 
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-19
BNC T
Adapter
AWG610 Arbitrary Waveform
Generator rear panel
Output
CH2 input
CH1 inputCH1 output
OscilloscopeĂ(TDS700)
SMA(Fe)ĆBNC(Ma)
adapter
50 W SMAĂcoaxial cable
50 W BNCĂcoaxial cable
50 W BNCĂcoaxial cable
Function Generator (AFG310)
Figure B-6: Triggered mode initial test hookup
4. Set the oscilloscope controls as follows:
Vertical . ........................ CH1andCH2
CH1 coupling . ............... DC
CH1 scale . ................. 0.5V/div
CH2 scale . ................. 2V/div
CH1 input impedance . ......... 50W
CH2 input impedance . ......... 1MW
Horizontal
Sweep . .................... 200ns/div
Trigger
Source . .................... CH1
Coupling .................... DC
Slope ...................... Positive
Level ...................... +100 mV
Mode . ..................... NORMAL