FG Mode
3-224
AWG610 Arbitrary Waveform Generator User Manual
The output parameter menu selections are the same for each waveform except
Pulse and DC. Pulse has one extra side menu item (Duty), and DC has only one
side menu item (Offset).
Output
parameters
Figure 3-75: Output parameters
Frequency
The frequency is set with a 4-digit number from 1.000 Hz to 260.0 MHz using
the SAMPLE RATE / SCALE knob, the numeric buttons or the general purpose
knob.The internal cut-off filter used is determined by the waveform type and the
frequency selected. The cut-off frequencies are as follows;
Table 3-54: Output frequency and filter cutĆoff frequency
Waveform type Output Frequency Filter Cut-off Frequency
Sine 1.000ĂHz to 260.0ĂKHz
260.1ĂKHz to 2.600ĂMHz
2.601ĂMHz to 8.000ĂMHz
8.001ĂMHz to 15.00ĂMHz
20 MHz
50 MHz
100 MHz
200 MHz
Output parameters