Tektronix AWG610 Portable Generator User Manual


 
The Graphical Waveform Editor
3-60
AWG610 Arbitrary Waveform Generator User Manual
Table 3-12: Set Pattern dialog box parameters (cont.)
Parameter Description
Pattern Specifies the pattern field value. Enter the pattern data by
using the 0 or 1 numeric buttons on the front panel or from an
attached keyboard.
Push the Clear Pattern side button to clear the pattern data
field.
Push the Import Pattern side button to insert the edit area
pattern data from the active window target data type into the
pattern field. You can then write the pattern data to a target
waveform type in the active window or another window.
Target Specifies the location where the generated data is created or
the source for imported pattern data. If you specify Data, the
pattern data replaces edit area waveform data; if Marker1 or
Marker2 is selected, the pattern data replaces the edit area
marker data. To import the pattern from the Target specified
here, use the Import Pattern side menu.
Do the following steps to specify a pattern:
1. Move the cursors to specify the edit area in which to replace the waveform
data with pattern data.
2. Push Operation (bottom)!Set Pattern... (pop-up)!OK (side).
The Set Pattern dialog box appears.
3. Select Data, Marker1, or Marker2 to specify the target data type to replace
with the pattern data.
4. Define the pattern using numeric buttons, or push the Import Pattern side
button to import the pattern data.
5. If necessary, you can change the pattern value by moving the cursor with the
a or ' button and then using numeric keys and the
key.
6. Push the OK side button to replace the waveform or marker data with the
specified pattern data.
The Import Pattern function lets you read waveform or pattern data from the
specified target data type of the active window and stores it in the pattern buffer.
You can then replace waveform or marker data with the pattern data in the
current window or another window. The Set Pattern dialog box converts all
waveform data greater than 0.5 volts to a one level if the waveform data is
analog T. All waveform data less than or equal to 0.5 volts is set to a zero level.