Appendix B: Performance Verification
B-52
AWG610 Arbitrary Waveform Generator User Manual
1/4 Clock FrequencyĂandĂ10 MHz Reference Input Tests
These procedures check the 10 MHz reference input function of the AWG610
Arbitrary Waveform Generator.
Equipment
required
Two 50ĂW BNCĂcoaxial cables, a frequency counter, and a function
generator.
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50 W BNC coaxial cable to connect the AWG610 Arbitrary Waveform
Generator 1/4 CLOCK OUT connector to the input A connector on the
frequency counter.
2. Use a 50 W BNC coaxial cable to connect the AWG610 Arbitrary Waveform
Generator 10 MHz REF IN connector to the function generator output
connector (see Figure B–28).
Function Generator (AFG310)
Frequency Counter
AWG610 Arbitrary Waveform
Generator rear panel
50 W BNCĂcoaxial cable
50 W BNCĂcoaxial cable
1/4ĂCLOCKĂOUT
10ĂMHzĂREFĂIN
Figure B-28: 1/4 Clock frequency and 10 MHz reference input initial test hookup
a. Set the frequency counter controls as follows:
INPUT A
Coupling . ................... AC
FUNCTION ...................... AFREQ
Gate time . ...................... 0.2s
Trigger Level . .................... 0V