Appendix B: Performance Verification
B-4
AWG610 Arbitrary Waveform Generator User Manual
Figure B-1: Diagnostic menu
Do the following to execute all of the AWG610 Arbitrary Waveform Generator
diagnostics automatically:
Push the Execute Diagnostic side button.
The internal diagnostics do an extensive verification of AWG610 Arbitrary
Waveform Generator functions. While this verification progresses, the screen
displays the clock icon. When finished, the resulting status appears on the
screen.
Verify that no failures are found and reported on-screen. If the diagnostics
terminates without error, Pass is displayed instead of the - - -. If a value is
displayed, meaning an error is detected, consult a qualified service technician for
further assistance.
Push any bottom or menu button (other than UTILITY) to exit the diagnostic
screen.
The instrument includes internal calibration routines that check electrical
characteristics such as offset, attenuations and filters. Perform calibration to
adjust internal calibration constants as necessary. This procedure describes how
to do the internal calibration.
Calibration