Tektronix AWG610 Portable Generator User Manual


 
Appendix B: Performance Verification
B-32
AWG610 Arbitrary Waveform Generator User Manual
2. Set the oscilloscope controls as follows:
Vertical . ........................ CH1
CH1 coupling . ............... DCifapplicable
CH1 scale . ................. 100mV/div
Horizontal
Sweep . .................... 500ps/div
Trigger
Source..................... CH1
Slope ......................
Level ...................... 0V
Mode . ..................... Auto
NOTE. The pulse rise time tests use the AWG610 Arbitrary Waveform Generators
control settings that have been used in the amplitude and DC offset tests. Do not
initialize the AWG610 Arbitrary Waveform Generator controls.
3. Load the PULSE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
4. Push SETUP (front-panel)!Run Mode (bottom)!Continuous (side).
The AWG610 Arbitrary Waveform Generator is set to the Continuous mode.
5. Change the AWG610 Arbitrary Waveform Generator controls as follows:
a. Push VERTICAL MENU (bottom)!Amplitude (side).
b. Push 0, ., 5 and ENTER keys in this order or turn the general pur-
pose knob to set the amplitude to 0.5 V.
c. Verify that the Direct DA out is set. If not, push Output (side)!Direct
(side).
6. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connectors are on.
7. Verify that the rise time of the pulse waveform displayed on the oscilloscope
is equal to or less than 400 ps.
8. Do the following substeps to change the connection to check the CH1
:
a. Push the CH1 output button.
b. Disconnect the cable from the CH1 output connector.