Tektronix AWG610 Portable Generator User Manual


 
Appendix B: Performance Verification
B-54
AWG610 Arbitrary Waveform Generator User Manual
1/4 Clock Output ĂLevelĂTests
These procedures check the AWG610 Arbitrary Waveform Generator clock
output signal.
Equipment
required
A 50ĂW BNC coaxial cable and an oscilloscope.
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use the 50 W BNC coaxial cable to connect the AWG610 Arbitrary
Waveform Generator 1/4 CLOCK OUT output to the oscilloscope input
connector (see Figure B–29).
AWG610 Arbitrary Waveform
Generator rear panel
OscilloscopeĂ(TDS700)
50 W BNCĂcoaxial cable
1/4ĂCLOCKĂOUT
Figure B-29: 1/4 Clock output level initial test hookup
2. Set the oscilloscope controls as follows:
Vertical . ........................ CH1
CH1 coupling . ............... DC
CH1scale .................. 0.5V/div
CH1 offset . ................. Ć1.3V
CH1 input impedance . ......... 1MW
Horizontal
Sweep . .................... 5ns/div
Trigger
Source . .................... CH1
Coupling .................... AC
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto