Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-45
Event Input and Enhanced Mode Tests
These procedures check the event input signals and enhanced mode operation.
NOTE. The event input check with strobe off and the strobe input check are
structured as a continuous test. After Check Event Input with Strobe Off, the
next test uses the connections and oscilloscope settings from the previous test.
Equipment
required
AĂ50ĂW SMAĂcoaxial cable,ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter an
oscilloscopeĂ(TDS700), and customĆmade ground closure. See Figure
B-3 for the connections.
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50W SMA coaxial cable and a SMA(Fe)-BNC(Fe) adapter to connect
the AWG610 Arbitrary Waveform Generator CH1 output connector to the
oscilloscope CH1 input connector (see Figure B–20).
OscilloscopeĂ(TDS700)
50 W SMAĂcoaxial cable
AWG610 Arbitrary Waveform Generator
SMA(Fe)ĆBNC(Ma)
adapter
Figure B-20: Event input and enhanced mode initial test hookup
2. Connect the ground closure to the EVENT IN connector on the AWG610
Arbitrary Waveform Generator rear panel.
3. Set the oscilloscope controls as follows:
Vertical......................... CH1
CH1 coupling . ............... DC
CH1scale .................. 0.2V/div
CH1 input impedance . ......... 50W
Check Event Input with
Strobe Off