Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-51
b. Close SW5 of the ground closure to generate an event signal on the
EVENT IN connector strobe pin.
c. Verify that the oscilloscope displays the DC waveform as shown in
Figure B–27.
Figure B-27: DC waveform output when the SW5 is closed
d. Open SW5 of the ground closure to degenerate the strobe signal on the
EVENT IN connector strobe pin.
e. Verify that the DC waveform is displayed on the oscilloscope.
f. Close SW5 of the ground closure again.
g. Verify that the oscilloscope displays the waveform as shown in Figure
B–26.
4. Push the CH1 OUT button to turn off the CH1 LED.
5. Disconnect the oscilloscope and ground closure.