Agilent Technologies 86100-90086 Sprinkler User Manual


 
1-44
Introduction
Commands Unavailable in Jitter Mode
RESPonse:TDRTDT
RESPonse:TDTDest
STIMulus
Timebase
Commands
MPOSition 23-2
Commands Unavailable in Jitter Mode
This section describes the commands that can generate errors when controlling the instru-
ment in Jitter mode. This can be due to the command or one of its arguments that are not
allowed in Jitter mode. Refer to the individual command reference for detailed information.
Refer to “New and Revised Commands” on page 1-42 for a list of commands that can be used
to control Jitter mode.
Waveform Files Waveform and Color Grade/Gray Scale files cannot be saved or loaded in Jitter mode. The
commands listed below produce a "Settings conflict" error when executed in Jitter Mode.
DISK:STORe 10-9
When used with sources other than SETup and JDMemory.
STORe:WAVeform 4-14
ACQuire:SWAVeform 6-6
LTESt:SWAVeform 15-8
MTESt:SWAVeform 17-13
Waveform
Queries
Only jitter database waveforms may be set or queried in Jitter mode. Using the following com-
mand produces the error, "Signal or trigger source selection is not available".
:WAVeform:DATA 25-4
Waveform
Memory Load/
Store
Waveforms cannot be saved into waveform memories in Jitter mode. All waveform memories
are turned off when entering Jitter mode. The commands listed below produce a "Settings
conflict" error when executed in Jitter mode.
WMEMory<N>:LOAD 26-2
WMEMory<N>:SAVE 26-3
DISK:LOAD 10-3
When used with sources other than SETup and JDMemory.