Agilent Technologies 86100-90086 Sprinkler User Manual


 
8-7
Channel Commands
UNITs
This command sets the TDR skew for the given channel. The TDR skew control moves the
TDR step relative to the trigger position. The control may be set from –100 to 100 percent of
the allowable range. This command is only applicable to TDR channels. This command is
enabled only if a stimulus is currently active and if the module has differential capability. <N>
represents the channel number and is an integer 1 to 4 followed by an optional A or B identi-
fying which of two possible channels in the slot is being referenced.
<percent> A number between –100 and 100, used to set the step position.
Example The following example sets the TDR skew for channel 1 to 20%.
10 OUTPUT 707;":CHANNEL1:TDRSKEW 20"
Query :CHANnel<N>:TDRSkew?
The query returns the current TDR skew setting for the specified channel.It returns the TDR
skew value in percent of allowable range from –100 to 100 percent. This command is only
applicable to TDR channels. The returned format is a real number.
Returned Format [:CHANnel<N>:TDRSkew] <value><NL>
UNITs
Command :CHANnel<N>:UNITs {VOLT | OHM |AMPere | REFLect | WATT | UNKNown}
This command sets the transducer units in Oscilloscope and Eye/Mask modes. In TDR/TDT
mode this command sets the channel units (VOLT, OHM, REFLect). <N> represents the
channel number and is an integer 1 to 4.
Query :CHANnel<N>:UNITs?
Returned Format [:CHANnel<N>:UNITs] {VOLT | OHM | REFLect | AMPere | WATT | UNKNown}<NL>
UNITs:ATTenuation
Command :CHANnel<N>:UNITs:ATTenuation <attenuation>
This command sets the transducer conversion factor. It provides the equivalent function of
the Transducer Conversion Factors Gain setting under the Setup menu’s Channel command.
This command is disabled for TDR channels and destinations channels for TDR/TDT mea-
surements. <N> represents the channel number and is an integer 1 to 4.
Query :CHANnel<N>:UNITs:ATTenuation?
Returned Format [:CHANnel<N>:UNITs:ATTenuation] <attenuation><NL>
UNITs:OFFSet
Command :CHANnel<N>:UNITs:OFFSet <offset>
This command sets the transducer offset. It provides the equivalent function of the Trans-
ducer Conversion Factors Offset setting under the Setup menu’s Channel command. This
command is disabled for TDR channels and destinations channels for TDR/TDT measure-
ments. <N> represents the channel number and is an integer 1 to 4.
Query :CHANnel<N>:UNITs:OFFSet?
Returned Format [:CHANnel<N>:UNITs:OFFSet] <offset><NL>