Agilent Technologies 86100-90086 Sprinkler User Manual


 
19-4
S-Parameter Commands
MAGGraph:VERTical:MAXimum
Sets the frequency span of the S-parameters graph.
Example 10 OUTPUT 707;":SPAR:MAGG:HOR:SPAN 5.0E+9"
Query :SPARameter:MAGGraph:HORizontal:SPAN?
Returned Format [:SPARameter:MAGGraph:HORizontal:SPAN] <span_freq><NL>
MAGGraph:VERTical:MAXimum
Command :SPARameter:MAGGraph:VERTical:MAXimum <vertical_max>
Sets the maximum amplitude (dB) of the S-parameters graph.
Example 10 OUTPUT 707;":SPAR:MAGG:VERT:MAX 5"
Query :SPARameter:MAGGraph:VERTical:MAXimum?
Returned Format [:SPARameter:MAGGraph:VERTical:MAXimum] <vertical_max><NL>
MAGGraph:VERTical:MINimum
Command :SPARameter:MAGGraph:VERTical:MINimum <vertical_min>
Sets the minimum amplitude (dB) of the S-parameters graph.
Example 10 OUTPUT 707;":SPAR:MAGG:VERT:MIN -30"
Query :SPARameter:MAGGraph:VERTical:MINimum?
Returned Format [:SPARameter:MAGGraph:VERTical:MINimum] <vertical_min><NL>
MARKer:X1STate
Command :SPARameter:MARKer:X1STate {ON | 1 | OFF | 0}
Turn on and off the X1 marker.
Example 10 OUTPUT 707;":SPAR:MARK:X1ST ON"
Query :SPARameter:MARKer:X1STate?
Returned Format [:SPARameter:MARKer:X1STate] {ON | 1 | OFF | 0}<NL>
MARKer:X2STate
Command :SPARameter:MARKer:X2STate {ON | 1 | OFF | 0}
Turn on and off the X2 marker.
Example 10 OUTPUT 707;":SPAR:MARK:X2ST ON"
Query :SPARameter:MARKer:X2STate?
Returned Format [:SPARameter:MARKer:X2STate] {ON | 1 | OFF | 0}<NL>
MARKer:X1Source
Command :SPARameter:MARKer:X1Source {CHANnel<N> | RESPonse<N> | WMEMory<N> | FUNCtion<N>}
Selects the source waveform of the X1 marker, if more than one waveform is displayed on the
graph.
Example 10 OUTPUT 707;":SPAR:MARK:X1S CHAN2"