Agilent Technologies 86100-90086 Sprinkler User Manual


 
24-6
Trigger Commands
SLOPe
SLOPe
Command :TRIGger:SLOPe {POSitive | NEGative}
This command specifies the slope of the edge on which to trigger.
Query :TRIGger:SLOPe?
The query returns the slope for the trigger.
Returned Format [:TRIGger:SLOPe] {POSitive | NEGative}<NL>
Example 10 OUTPUT 707; ":TRIGger:SLOPe POSitive"
SOURce
Command :TRIGger:SOURce {FPANel | FRUN | LMODule | RMODule}
This command selects the trigger input. Front Panel, Left Module, and Right Module are
inputs from the front panel of the instrument. Free Run is internally generated, and is not
affected by the settings of gates, level, slope, bandwidth, or hysteresis.
<trigger> Front PANel, Left MODule, and Right MODule are inputs on the front of the instrument. Fre-
eRUN is internally generated and is unaffected by the settings for gated, level, slope, band-
width or hysteresis.
Query :TRIGger:SOURce?
The query returns the current trigger source of the current mode.
Returned Format [:TRIGger:SOURce] <trigger><NL>