Agilent Technologies 86100-90086 Sprinkler User Manual


 
11-5
Display Commands
JITTer:LAYout
JITTer:LAYout
Command :DISPlay:JITTer:LAYout {SINGle|SPLit|QUAD}
This command sets the number of graphs displayed when in jitter mode. SINGle specified one
graph, SPLit specifies two graphs and QUAD specifies four graphs.
Restrictions Jitter mode (86100C instruments). Software revision A.04.00 and above with Option 100 or
software revision A.04.10 and above with Option 200.
Example 10 OUTPUT 707; ":DISPlay:JITTer:LAYout SPLit"
Query :DISPlay:JITTer: LAYout?
Returned Format Returns the current setting for jitter mode graph layout.
JITTer:PJWFrequency
Command :DISPlay:JITTer:PJWFrequency <frequency>
For the PJ Waveform graph, sets or queries the frequency plotted on the graph. The com-
mand, :DISPlay:JITTer:PJWTracking, must be set to “off” before issuing the PJWFrequency
command or query.
Restrictions Jitter mode. Software revision A.04.10 and above (86100C instruments). Option 200,
Enhanced Jitter Analysis Software.
Query :DISPlay:JITTer:PJWFrequency?
Returned Format [:DISPlay:JITTer:PJWFrequency] <frequency><NL>
Example 10 OUTPUT 707;”:DISPlay:JITTer:PJWFrequency 10E+6”
JITTer:PJWTracking
Command :DISPlay:JITTer:PJWTracking {{ON | 1}|{OFF | 0}}
For the PJ Waveform graph, sets or queries the option for automatically tracking the fre-
quency component with the greatest magnitude.
Restrictions Jitter mode. Software revision A.04.10 and above (86100C instruments). Option 200,
Enhanced Jitter Analysis Software.
Query :DISPlay:JITTer:PJWTracking?
Returned Format [:DISPlay:JITTer:PJWTracking] {{ON | 1}|{OFF | 0}}<NL>
Example 10 OUTPUT 707;”:DISPlay:JITTer:PJWTracking ON”
JITTer:SHADe
Command :DISPlay:JITTer:SHADe {{ON | 1}|{OFF | 0}}
Shows or removes the display of the jitter shade. The shade is the drop-down screen that is
used to display the jitter graphs. Because showing the shade takes some time, use this com-
mand to reduce measurement times in situations where testing would continually open and
hide the jitter shade.
Restrictions Jitter mode. Software revision A.06.00 and above (86100C instruments). Option 200,
Enhanced Jitter Analysis Software.