Tektronix AWG510 Portable Generator User Manual


 
Performance Verification
AWG510 & AWG520 Service Manual
4-27
6. End procedure: Turn off the function generator output. Retain the test
hookup.
Equipment
required
Three 50Ăcoaxial cables, an adapter (BNC T male to 2 females), a
function generator, and an oscilloscope.
Prerequisites The AWG500-Series Waveform Generator meets the prerequisites
listed on page 4-13.
1. Use the test hookup from previous check.
2. Set the test equipment controls:
a. Set the oscilloscope controls:
Vertical......................... CH1andCH2
CH1 and CH2 coupling . ........ DC
CH1scale .................. 0.5V/div
CH2scale .................. 2V/div
CH1 input impedance . ......... 50W
CH2 input impedance . ......... 1MW
Horizontal
Sweep . .................... 1ms/div
Trigger
Source . .................... CH1
Coupling .................... AC
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto
b. Set the function generator (AFG310) controls:
Function ........................ Square
Mode .......................... Continuous
Parameters ......................
Frequency .................. 10kHz
Amplitude . .................. 2.0V(4Vinto open circuit)
Offset . ..................... 1.0V(2Vinto open circuit)
Output . ........................ Off
3. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
Check Gated Mode