Tektronix AWG510 Portable Generator User Manual


 
Performance Verification
AWG510 & AWG520 Service Manual
4-31
b. Select the file: Load the MODE.WFM as referring to the procedures on
page 4–15.
c. Set clock frequency:
H Push HORIZONTAL MENU (front–panel)!Clock (side).
H Enter numeric value of 10M: Push 1, 0 and M (SHIFT+7) keys in
this order.
3. Turn on the AWG500–Series Waveform Generator output: Push the RUN
button so that the LEDs above the RUN button lights.
4. Check clock frequency accuracy: Check that the frequency counter reading is
10 MHz ±10 Hz (1 ppm).
5. End procedure: Disconnect the frequency counter.