Tektronix AWG510 Portable Generator User Manual


 
Specifications
1-6
AWG510 & AWG520 Service Manual
Table 1-6: Main output (Cont.)
Characteristics
3
PV reference pageDescription
Normal out
Output voltage -2.0 V to +2.0V, into a 50 W load
Amplitude
Range 20 mV
p-p
to2V
p-p
,intoa50W load
Resolution 1mV
* DC accuracy
"(1.5 % of amplitude + 2 mV), offset: 0 V
Page 4-32
Offset
Range -1.000 V to 1.000 V, into a 50 W load
Resolution 1mV
* Offset accuracy
"(1 % of offset + 10 mV),
20 mV amplitude (waveform data: 0)
Page 4-32
Reverse power protection Up to 0.4 W
* Pulse response (Waveform data: -1 and 1, offset: 0 V, and filter: through) Page 4-40
Rise time (10 % to 90 %)
x 2.5 ns (amplitude u 1.0 V
p-p
)
x 1.5 ns (amplitude x 1.0 V
p-p
)
Fall time (10 % to 90 %)
x 2.5 ns (amplitude u 1.0 V
p-p
)
x 1.7 ns (amplitude x 1.0 V
p-p
)
* Aberration (500 MHz Bandwidth)
"10 % (amplitude u 1.0 V
p-p
)
"7 % (amplitude x 1.0 V
p-p
)
Page 4-40
* Flatness
"3 % (after 50 ns from rise and fall edges)
Page 4-40
Small signal bandwidth, Typical 300 MHz (-3 dB, amplitude fixed to 0.5 V)
* Sinewave characteristics
(CH1 and CH2)
(Clock: 1 GHz, waveform points: 32, frequency: 31.25 MHz,
amplitude: 1.0 V, offset: 0 V, filter: through)
Page 4-43
Harmonics
x -50 dBc (DC to 400 MHz)
Noise
x -53 dBc (DC to 400 MHz)
Phase Noise
x -90 dBc / Hz (10 kHz offset)