Tektronix AWG510 Portable Generator User Manual


 
Performance Verification
4-72
AWG510 & AWG520 Service Manual
Digital Data Output Tests (Option 03 Only)
These procedures check the AWG500–Series Waveform Generator digital data
output level accuracy and skew.
Equipment
required
A 50ĂSMBĆtoĆBNC coaxial cable, and an oscilloscope.
Prerequisites The instrument must meet the prerequisites listed on page 4-13.
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG500–Series Waveform
Generator DIGITAL DATA OUT:D0 output through a 50 SMB-to-
BNC coaxial cable to the oscilloscope input connector (see Figure 4–34).
Connect the SMB
connector end to DIGITAL
DATA OUT:D0 connector
Oscilloscope
AWG500 Series Waveform
Generator rear panel
50 W SMBĆtoĆBNC coaxial cable
Figure 4-34: Digital data output level initial test hookup
b. Set the oscilloscope controls:
Vertical . ........................ CH1
CH1 coupling . ............... DC
CH1 scale . ................. 1V/div
CH1offset .................. 0V
CH1 input impedance . ......... 50W
Horizontal
Sweep ..................... 1ms/div
Trigger
Source . .................... CH1
Coupling .................... AC
Check Output Levels