Tektronix AWG510 Portable Generator User Manual


 
Specifications
1-16
AWG510 & AWG520 Service Manual
Environmental Specification
Table 1-21: Installation requirement
Characteristics Description
Heat dissipation
Maximum power 600 W (maximum line current: 8 Arms, at 50 Hz)
Surge Current 30 A (25 _C) peak for equal to or less than 5 line cycles,
after the instrument has been turned off for at least 30s
Cooling clearance Bottom 2 cm (0.8 inches)
NOTE: The feet on the bottom provides the required
clearance when set on a flat surface.
Sides 15 cm (6 inches)
Rear 7.5 cm (3 inches)
Table 1-22: Environmental
Characteristics Description
Atmospherics
Temperature
Operating +10 _Cto+40_C
NonĆoperating -20 _Cto+60_C
Relative humidity
Operating 20 % to 80 % (no condensation)
Maximum wetĆbulb temperature 29.4 _C
NonĆoperating 5 % to 90 % (no condensation)
Maximum wetĆbulb temperature 40.0 _C
Altitude
Operating Up to 3 km (10 000 ft)
Maximum operating temperature decreases 1 _C each 300 m (1 000 ft)
above 1.5 km (5 000 ft)
NonĆoperating Up to 12 km (40 000 ft)
Dynamics
Random vibration
Operating 0.27 g rms, from 5 Hz to 500 Hz, 10 minutes
NonĆoperating 2.28 g rms, from 5 Hz to 500 Hz, 10 minutes
Shock
NonĆoperating 294 m/s
2