Tektronix AWG510 Portable Generator User Manual


 
Performance Verification
4-36
AWG510 & AWG520 Service Manual
Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out)
These procedures check the accuracy of the AWG500–Series Waveform
Generator direct waveform outputs; amplitude and offset.
Equipment
required
A 50Ăcoaxial cable, a 50Ăprecision terminator, a BNC (female)ĆtoĆ
dual banana adapter, and a digital multimeter (DMM).
Prerequisites The AWG500-Series Waveform Generator meets the prerequisites
listed on page 4-13.
1. Install the test hookup and set test equipment controls:
a. Hook up DMM: Connect the AWG500–Series Waveform Generator CH1
output through a 50 BNC coaxial cable, a 50 precision terminator,
and an adapter (BNC-to-dual banana) to the DMM input connector (see
Figure 4–11).
AWG500 Series Waveform Generator
DMM
50 W coaxial cable
Dual Banana
Adapter
50 W
Terminator
Figure 4-11: Direct DA out amplitude accuracy initial test hookup
b. Set the DMM controls:
Mode . ......................... VDC
Range ......................... 2V
Input . ......................... Front
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)!
System (bottom)!Factory Reset (side)!OK (side).
Check Amplitude
and
DC Offset