S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb
Error
−
260 Expression error
Indicates that an expression program data element related error occurred.
Error
−
261 Math error in expression
Indicates that a syntactically legal expression program data element could not be executed
due to a math error. For example, a divide-by-zero was attempted.
Error
−
270 Macro error
Indicates that a macro-related execution error occurred.
Error
−
271 Macro syntax error
Indicates that a syntactically legal macro program data sequence, according to IEEE
488.2, 10.7.2, could not be executed due to a syntax error within the macro definition (see
IEEE 488.2, 10.7.6.3).
Error
−
272 Macro execution error
Indicates that a syntactically legal macro program data sequence could not be executed
due to some error in the macro definition (see IEEE 488.2, 10.7.6.3).
Error
−
273 Illegal macro label
Indicates that the macro label defined in the *DMC command was a legal string syntax,
but could not be accepted by the device (see IEEE 488.2, 10.7.3 and 10.7.6.2). For
example, the label was too long, the same as a common command header, or contained
invalid header syntax.
Error
−
274 Macro parameter error
Indicates that the macro definition improperly used a macro parameter placeholder (see
IEEE 488.2, 10.7.3).
Error −275 Macro definition too long
Indicates that a syntactically legal macro program data sequence could not be executed
because the string of block contents were too long for the device to handle (see IEEE