Equivalent Front-Panel Key Commands
To Set the Meter High End and Low End Point Display Units.
Use the
:METer:HEND:DUNits and :MEter:LEND:DUNits commands to set the analog
bar-graph meter high end point and low end point Display Units. Refer to
“Display
Units (DUNits)” on page 75
for description of Display Units.
Syntax
:METer:HEND:DUNits <disp_units>
:METer:HEND:DUNits <disp_units>
Example
OUTPUT 714;"MEAS:RFR:POW:MET:HEND:DUN DBM"
OUTPUT 714;"MEAS:RFR:POW:MET:LEND:DUN DBM"
This sets the high end point and low end point display units of the analog
bar-graph meter for the TX Power measurement to DBM.
NOTE: When querying the METER Data Function through GPIB, the Test Set always
returns numeric values in Attribute Units, regardless of the current Display Units
or GPIB Units settings. Numeric values are expressed in scientific notation.
To Query the Meter High End and Low End Point Display Units.
Use the
:METer:HEND:DUNits? and :MEter:LEND:DUNits? commands to query the
analog bar-graph meter high end point and low end point Display Units. Refer to
“Display Units (DUNits)” on page 75 for description of Display Units.
Syntax
:METer:HEND:DUNits?
:METer:LEND:DUNits?
Example
OUTPUT 714;"MEAS:RFR:POW:MET:HEND:DUN?
OUTPUT 714;"MEAS:RFR:POW:MET:LEND:DUN?"
ENTER 714;Met_hidisp_unit$
ENTER 714;Met_lodisp_unit$
This queries the high end point and low end point display units of the
analog bar-graph meter for the TX Power measurement.
NOTE: When querying the METER Data Function through GPIB, the Test Set always
returns numeric values in Attribute Units, regardless of the current Display Units
or GPIB Units settings. Numeric values are expressed in scientific notation.