Chapter 7, IBASIC Controller
Interfacing to the IBASIC Controller using Serial Ports
6. Press DONE to return to the Config screen.
7. Press TERMINAL CONFIG.
Terminal Id: 2392A
LocalEcho: OFF
CapsLock: OFF
Start Col: 01
Bell: ON
XmitFnctn(A): NO
SPOW(B): NO
InhEolWrp(C): NO
Line/Page(D): LINE
InhHndShk(G): NO
Inh DC2(H): NO
Esc Xfer(N): YES
ASCII 8 Bits: YES
Fld Separator: down arrow or US
BlkTerminator: up arrow or RS
ReturnDef: musical note or CR
Copy: Fields
Type Ahead: NO
Row Size: 160
Host Prompt Character: left arrow or D1
Horiz. Scrolling Increment: 08
8. Press DONE to return to the Config screen.
9. Press DONE to return to the Terminal screen.
10. Press MAIN to return to the Main screen.
11. Press EXIT ADVLINK to exit.
Terminal Configuration
Use the cable information in Table 39 on page 362 and Figure 25 on page 363 for
connecting to an external terminal. Terminals typically have a DB-25 (male)
connector. Set the terminal for DEC VT-100 ANSI emulation. Many ASCII
terminals will also function properly.
To set up the terminal, use the field settings found in the Agilent AdvanceLink
terminal emulator section found earlier in this chapter. As a minimum, make sure
the terminal’s basic setup information matches the fields on the Test Set’s I/O
CONFIGURE screen (refer to
Table 41 on page 366 for recommended settings).