Agilent Technologies A.18.00 Weather Radio User Manual


 
222
S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec
IEEE 488.2 Common Commands
*RCL
(Recall Instrument
State)
The *RCL command restores the state of the Test Set from a file previously stored
in battery-backed internal memory, on a memory card, on a RAM disk, or on an
external disk. The *RCL command is followed by a decimal number in the range
of 0 to 99 which indicates which Test Set SAVE/RECALL file to recall. The mass
storage location for SAVE/RECALL files is selected using the
SAVE/RECALL
field on the I/O CONFIGURE screen.
The *RCL command cannot be used to recall files with names which contain
non-numeric characters or a decimal number greater than 99. To recall
SAVE/RECALL files saved with names which contain non-numeric characters or
a decimal number greater than 99, use the REG:RECall filename command (see
RECALL in the “Equivalent Front-Panel Key Commands” section of chapter 4 of
the Agilent 8920B Programmer’s Guide).
*SAV
(Save Instrument
State)
The *SAV command saves the present state of the Test Set into a file in battery-
backed internal memory, on a memory card, on a RAM disk, or on an external
disk. The *SAV command is followed by a decimal number in the range of 0 to 99
which indicates the name of the stored SAVE/RECALL file. The mass storage
location for SAVE/RECALL files is selected using the
SAVE/RECALL field on
the I/O CONFIGURE screen.
The *SAV command cannot be used to save the present state of the Test Set to a
file with a name which contains non-numeric characters or a decimal number
greater than 99. To save the present state of the Test Set to a file with a name
which contains non-numeric characters or a decimal number greater than 99, use
the REG:SAVE filename command (see SAVE in the “Equivalent Front-Panel
Key Commands” section of chapter 4 of the Agilent 8920B Programmer’s Guide).