48
S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb
Chapter 1, Using GPIB
Remote Operation
Remote Capabilities
Conformance to the IEEE 488.1-1987 Standard
For all IEEE 488.1 functions implemented, the Test Set adheres to the rules and
procedures as outlined in that Standard.
Conformance to the IEEE 488.2-1987 Standard
For all IEEE 488.2 functions implemented, the Test Set adheres to the rules and
procedures as outlined in that Standard with the exception of the *OPC Common
Command. Refer to the *OPC Common Command description.
IEEE 488.1 Interface Functions
The interface functions that the Test Set implements are listed in Table 5.
Table 5 Test Set IEEE 488.1 Interface Function Capabilities
Function Capability
Talker T6: No Talk Only Mode
Extended Talker T0: No Extended Talker Capability
Listener L4: No Listen Only Mode
Extended Listener LE0: No Extended Listener Capability
Source Handshake SH1: Complete Capability
Acceptor Handshake AH1: Complete Capability
Remote/Local RL1: Complete Capability
Service Request SR1: Complete Capability
Parallel Poll PP0: No Parallel Poll Capability
Device Clear DC1: Complete Capability
Device Trigger DT1: Complete Capability
Controller C1: System Controller
C3: Send REN
C4: Respond to SRQ
C11:No Pass Control to Self, No Parallel Poll
Drivers E2: Tri-State Drivers