198
S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec
Equivalent Front-Panel Key Commands
To Set A Measurement Reference Point.
Use the :REFerence:VALue commands to
set a measurement reference point.
Syntax
:REFerence:VALue <real number
Example
OUTPUT 714;"MEAS:RFR:POW:REF:VAL 20"
This sets the measurement reference point for the TX Power
measurement to 20 watts.
NOTE: When setting a measurement reference point, any non–Attribute Unit “unit-of-
measure” must be specified in the command string, otherwise the current Attribute
Unit is assumed by the Test Set. Refer to “Attribute Units (AUNits)” on page 81.
To Query A Measurement Reference Point. Use the :REFerence:VALue?
commands to query a measurement reference point.
Syntax
:REFerence:VALue?
Example
OUTPUT 714;"MEAS:RFR:POW:REF:VAL?"
ENTER 714;Ref_val
This queries the measurement reference point for the TX Power
measurement.
NOTE: When querying a measurement reference point through GPIB, the Test Set always returns
numeric values in Attribute Units, regardless of the current Display Units or GPIB Units
settings. Numeric values are expressed in scientific notation. Refer to “Attribute Units
(AUNits)” on page 81.