Chapter 8, Programming the Call Processing Subsystem
Programming the ANALOG MEAS Screen
AFGen1 Freq
This field sets the output frequency of Audio Frequency Generator #1. See “AF
Analyzer” on page 97
for programming command syntax.
AFGen1 To
This field has two subfields:
• the upper subfield sets the destination port for Audio Frequency Generator #1
• FM = RF Generator FM modulator
• AM = RF Generator AM modulator
• Audio Out = AUDIO OUT connector on front panel of Test Set
• the lower subfield sets the:
• FM modulation deviation if upper subfield set to FM
• AM modulation depth if upper subfield set to AM
• amplitude of audio signal (volts RMS) at the AUDIO OUT connector if upper sub-
field set to Audio Out
For testing mobile stations the upper field is normally set to FM and the lower field
set to the desired FM deviation in kHz.
See “AF Generator 1” on page 100 for
programming command syntax.
Amplitude
This field sets the output power of the Test Sets’s transmitter (that is, the output
power of the Test Set’s RF Generator).
See “RF Generator” on page 163 for
programming command syntax.
De-Emphasis
This field is used to select or bypass the 750 uSec de-emphasis filter network used
to condition the audio signal before being analyzed by the Audio Frequency
Analyzer.
See “AF Analyzer” on page 97 for programming command syntax.
Detector
This field is used to select the type of detector used to measure the amplitude of
the audio signal being analyzed by the Audio Frequency Analyzer.
See “AF
Analyzer” on page 97
for programming command syntax.