80
Chapter 5: Troubleshooting
Troubleshooting the Agilent 1680A,AD-series
Troubleshooting the Agilent 1680A,AD-series
Apply power. Observe
the instrument
front panel.
Did the front
panel indicators
flash?
Yes
Launch the
Agilent Logic Analyzer
application software.
Is the problem
still present?
Do the self test.
Refer to Chapter 5.
Yes
No
Replace the
acquisition
board.
No
Do the self
tests pass?
Yes
No
Yes
Is the
application
Online?
Yes
No
Does the
application software
launch?
Does the
instrument finish
booting?
No
Does the
display light
up?
No
Yes
No
Yes
No
Yes
START
DONE
2
3
4
5
6A
6B
7
Are all
instrument power
supply fans
running?