Chapter 3: Testing Performance
To test the single-clock, multiple-edge, state acquisition
2 Check the data pulse width. Using the oscilloscope, verify that the data
pulse width is 3.000 ns, +0 ps or 100 ps:
a In the oscilloscope Timebase menu, select Scale: 1.000 ns/div.
b In the oscilloscope Timebase menu, select Position. Using the
oscilloscope knob, position the data waveform so that the waveform is
centered on the screen.
c On the oscilloscope, select [Shift] + width: channel 1, then select
[Enter] to display the data signal pulse width (+ width(1)).
d If the pulse width is outside the limits, adjust the pulse generator
channel 2 width until the pulse width is within limits.
Check the setup/hold with single clock, multiple clock
edges
The following setup/hold combinations will be tested.
Setup/Hold Combinations
Te s t
Combination
Setup/Hold
Times
Setup/Hold
Window
Sample Position
(in middle of Window)
1 5.0/-2.0 ns 3.0 ns -3.5 ns
2 -1.5/4.50 ns 3.0 ns +3.0 ns