133
Chapter 6: Replacing Assemblies
1690A,AD-series disassembly/assembly
1690A,AD-series disassembly/assembly
Prepare the instrument for disassembly
Do this procedure before doing any disassembly procedure on the instrument.
1 Close the Agilent Logic Analyzer application software.
2 Disconnect the logic analyzer from the host PC.
3 Remove power and disconnect the power cord.
4 Move the instrument to a static safe work environment.
To remove the chassis from the sleeve
Before disassembling the instrument, it must be turned off and placed in a static
safe work environment. If you haven't already done so, do the previous procedure
"Prepare the instrument for disassembly."
1 Using a Torx T-15 screwdriver, remove the two screws that secure the
handle to the side of the instrument and lift off the handle.
2 Using a Torx T-10 screwdriver, remove five screws that secure the logic
analyzer cables (and spacers, if installed) to the front panel of the logic
analyzer.
3 Disconnect the logic analyzer cables from the front panel. Remove the
logic analyzer cables (and spacers, if installed) from the logic analyzer.
4 Using a Torx T-10 screwdriver, remove thirteen screws that secure the
cover to the chassis.
5 With the logic analyzer upright, slide the chassis out of the cover.