45
Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
then select [Enter] to display the setup time (∆ Time(1)-(2)).
c Adjust the pulse generator channel 1 Delay until the pulses are aligned
according to the setup time of the setup/hold combination selected,
+0.0 ps or –100 ps.
9 Select the clock to be tested:
The following clock configurations will be used in steps 9, 10 and 11.
a In the Analyzer Setup dialog, click the Sampling tab.
b In the Sampling tab, click the Master button for the first clock to be
tested (Clk 1) and select Falling Edge.