56
Chapter 3: Testing Performance
To test the multiple-clock state acquisition
have been configured with Falling Edge.
d Click OK to close the Analyzer Setup dialog.
9 Verify the test data:
a Click the Run icon.
b If you have not already done so, do “Set up the Markers:” on page 34.
c If the "can't find 4096 occurence(s)" message does not appear, the test
passes.
The test passes when the logic analyzer finds all occurances of the patterns
programmed into the Markers. If the test passes, record a "Pass" in the
performance test record under single-clock single-edge next to the clock and
edge being tested.
10 Test the next setup/hold combination:
a Click the Bus/Signal Setup icon.
b Disconnect the clock just tested from the pulse generator.
c Repeat steps 1 through 10 for the next setup/hold combination listed in
step 1 in page 52.
When aligning the data and clock waveforms using the oscilloscope, align the
waveforms according to the setup time of the setup/hold combination being
tested, +0.0 ps or 100 ps.
Test the next channels (1680/81A, AD and 1690/91A, AD)
Connect the next combination of data channels and clock channels, then repeat
the previous test.
Start with “Connect and configure the logic analyzer” on page 49, connect the
next combination, then continue through the complete test.