Chapter 2: Preparing for Use
Power Requirements
The logic analyzer requires a power source of either 115 Vac or 230 Vac, –22 % to
+10 %, single phase, 48 to 66 Hz, CAT II pollution degree 2, 140/400 Watts
nominal maximum power (1680A/AD-series), and 76/200 Watts nominal
maximum power (1690A/AD-series).
Operating Environment
The operating environment is listed in chapter 1. The logic analyzer will operate
at all specifications within the temperature and humidity range given in chapter
1. However, reliability is enhanced when operating the logic analyzer within the
following ranges:
Temperature: +20° C to +35° C (+68° F to +95° F)
Humidity: 20% to 80% noncondensing
Note the recommended noncondensing humidity. Condensation within the
instrument can cause poor operation or malfunction. Provide protection against
internal condensation.
Storage
Store or ship the logic analyzer in environments within the following limits:
Temperature: -40° C to +75° C
Humidity: Up to 90% at 65° C
Altitude: Up to 15,300 meters (50,000 feet)
Protect the logic analyzer from temperature extremes which cause condensation
on the instrument.
To inspect the logic analyzer
1 Inspect the shipping container for damage.
If the shipping container or cushioning material is damaged, keep them until you
have checked the contents of the shipment and checked the instrument
mechanically and electrically.
2 Check the supplied accessories.
Accessories supplied with the logic analyzer are listed in “Accessories” on
page 10.