Agilent Technologies 1680 Weather Radio User Manual


 
62
Chapter 3: Testing Performance
To test the single-clock, multiple-edge, state acquisition
1 Using the Delay mode of the pulse generator channel 2, position the pulses
according to the setup time of the setup/hold combination selected,
+0.0 ps or –100 ps:
a On the Oscilloscope, select [Define meas] Define Time - Stop edge:
rising.
b In the oscilloscope timebase menu, select Position. Using the
oscilloscope knob, position the falling edge of the data waveform so that
it is near the center of the display.
c On the oscilloscope, select [Shift] Time. Select Start src: channel 1,
then select [Enter] to display the setup time ( Time(1)-(2)).
d Adjust the pulse generator channel 2 Delay until the pulses are aligned
according to the setup time of the setup/hold combination selected,
+0.0 ps or –100 ps.
2 Select the logic analyzer sample positions:
a Click the Sampling Setup icon. The Analyzer Setup dialog opens with
the Sampling tab displayed.
b Click Sample Positions....
c In the Sample Positions dialog, drag the blue bar for My Bus 1 to the
sample position of the first setup/hold combination to be tested (or
enter the value in the signal fields).