Chapter 5: Troubleshooting
Troubleshooting the Agilent 1690A,AD-series
Device Manager
Use Device Manager to see if the Agilent Logic Analyzer device has been properly
installed. The Device Manager should include an entry "Logic Analyzers" with the
device "Agilent Technologies 1680/1690 Series Analyzer".
If the logic analyzer device is not listed, then uninstall and reinstall the Agilent
Logic Analyzer application software. If the device still does not appear, then there
is a problem with the Windows XP Professional operating system. Consult the
documentation for the operating system to determine why the device is not being
installed and run.
To test the power supply voltages
Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace
covers and assemblies.
This procedure will not expose any problems related to load regulation; however,
it will show most failure modes to over 95% confidence,
WARNING: Hazardous voltages exist on the power supply. This procedure is to be
performed by service-trained personnel aware of the hazards involved, such as
fire and electrical shock.
1 Close the Agilent Logic Analyzer application and all other applications
running on the logic analyzer.
2 Shut down the instrument.