Chapter 3: Testing Performance
To set up the logic analyzer for the state mode tests
h Repeat steps b through f to configure marker M2 using the following
pattern according to the logic analyzer being tested.
1680A,AD, 1690A,AD - "55"
1681A,AD, 1691A,AD - "15"
1682A,AD, 1692A,AD - "55"
1683A,AD, 1693A,AD - "5"
i Click OK to close the Value dialog.
j In the Listing Viewer Properties dialog, select Beginning Of Data in the
“from” field.
The logic analyzer markers are now configured to verify the test data. If the error
message "can't find 4096 occurence(s)" does not appear, the test passes. Click
OK to close the Listing ViewerProperties dialog
When the above error message appears, one or more samples of test data is
incorrect. When this happens, check the following and attempt the test again:
All cables are properly connected.
Configuration of each test equipment is correct.
Logic analyzer is properly set up according to the test procedure.