Chapter 3: Testing Performance
To test the single-clock, multiple-edge, state acquisition
To test the single-clock, multiple-edge, state acquisition
Testing the single-clock, multiple-edge, state acquisition verifies the performance
of the following specifications:
Minimum master-to-master clock time.
Maximum state acquisition speed.
Setup/Hold time for single-clock, multiple-edge, state acquisition.
This test checks two combinations of data using a multiple-edge single clock at
two selected setup/hold times.
Equipment Required
Set up the equipment
1 If you have not already done so, do the following procedures:
“To set up the test equipment and the logic analyzer” on page 23
“To set up the logic analyzer for the state mode tests” on page 33
2 Modify the following pulse generator settings:
Period: 10.000 ns
Channel 2: Width 3.000 ns
Channel 2: Pulse ÷1
Channel 1: Pulse
Channel 1: Width 5.000 ns
Equipment Critical Specifications Recommended Model/Part
Pulse Generator 200 MHz 3.0 ns pulse width, < 600 ps rise time 8133A option 003
Digitizing Oscilloscope 6 GHz bandwidth, < 58 ps rise time 54750A w/ 54751A
Adapter SMA(m)-BNC(f) 1250-1200
SMA Coax Cable (Qty 3) 18 GHz bandwidth 8120-4948
Coupler BNC(m)(m) 1250-0216
BNC Test Connector,