Agilent Technologies 1680 Weather Radio User Manual


 
46
Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
c Click the Master buttons for the remaining clocks and select Don't Care
to turn off the other clocks.
d Connect the clock to be tested to the pulse generator channel 1 output.
e Click OK to close the Analyzer Setup dialog.
10 Verify the test data:
a Click the Run icon.
b If you have not already done so, do “Set up the Markers:” on page 34.
c If the "can't find 4096 occurence(s)" message does not appear, the test
passes.
The test passes when the logic analyzer finds all occurances of the patterns
programmed into the Markers. If the test passes, record a "Pass" in the
performance test record under single-clock single-edge next to the clock and
edge being tested.
11 Test the next clock:
a Click the Sampling Setup icon.
b Disconnect the clock just tested from the pulse generator.
c Repeat steps 9, 10, and 11 for the next clock configuration listed in step
9 until all listed clock combinations have been tested.
12 Test the next setup/hold combination:
a Click the Bus/Signal Setup icon.
b Disconnect the clock just tested from the pulse generator.
c Repeat steps 1 through 12 for the next setup/hold combination listed on
page 41.