Agilent Technologies 1680 Weather Radio User Manual


 
55
Chapter 3: Testing Performance
To test the multiple-clock state acquisition
6 Enable the pulse generator channel 1 COMP (with the LED on).
7 Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup/hold combination selected, +0.0 ps or –100 ps:
a On the Oscilloscope, select [Define meas] Define Time - Stop edge:
falling.
b On the oscilloscope, select [Shift] Time. Select Start src: channel 1,
then select [Enter] to display the setup time ( Time(1)-(2)).
c Adjust the pulse generator channel 1 Delay until the pulses are aligned
according to the setup time of the setup/hold combination selected,
+0.0 ps or –100 ps.
8 Select the clocks to be tested:
a Click the Sampling Setup icon. The Analyzer Setup dialog opens.
b In the Sampling tab, click the Master button for one of the clocks and
select Falling Edge.
c Repeat the above steps for each of the remaining clocks until all clocks