Tektronix MTG100, MTG300 Portable Generator User Manual


 
Performance Verification
MTG100 & MTG300 MPEG Generator Service Manual
4Ć9
7. Disconnect the BNC cable from the 75 W to 50 W minimum loss attenuator
on the oscilloscope’s input, and then connect the BNC cable to the SSI IN
connector on the MPEG test system’s Real Time Analyzer as shown in
Figure 4-6.
8. Use the parallel interface cable to connect the SSI parallel output to the
Real Time Analyzer parallel input on the MPEG test system rear panel as
shown in Figure 4-6.
9. Double-click the Real Time Analyzer icon in the MPEG test system’s
Program Group window to start the application.
10. In the Real Time Analyzer window on the MPEG test system, select
Settings... from the Configuration menu to open the Settings window.
11. In the Settings window, select Hardware Configuration to open the
Hardware Configuration window.
12. In the Hardware Configuration window, click the Restore Defaults button,
click the Apply button, and then select DVB-PI SPI (LVDS//) or ECL/ as
the Input Type.
13. In the Settings window, select ETR290 to open the ETR290 configuration
window.
14. In the ETR290 window, click the Restore Defaults button, click the Apply
button, and then click OK to close the Settings window.
15. In the Real Time Analyzer window, click the Start Analysis toolbar button.
16. Verify that the Hierarchic view is displayed on the screen and that no error
messages appear. In addition, verify that the transmission rate is 40 Mbps
and that the transport stream packet size is 188 bytes.
MPEG generator rear panel
75 W BNC Cable
Parallel interface cable
MPEG test system rear panel
(MTS200 or MTS215 Opt. SS)
Figure 4Ć6: Second equipment connections for verifying the SSI output