Agilent Technologies E4400-90333 Portable Generator User Manual


 
64 ErrorMessages
201 to 702: Device-Specific Errors ESG Family Signal Generators
616 Configuration Error; Attenuator configuration does not match
actual installed attenuator type.
A mismatch was found between the configured and detected attenuator
types. The instrument may not be fully functional. Report this error to the
nearest Agilent Technologies sales or service office.
Configuration Error; Option reconfiguration failed.
As a part of option reconfiguration, the instrument reboots. If the
reconfiguration failed, it is reported when the instrument completes the
reboot. Try repeating the option reconfiguration. Report this error to the
nearest Agilent Technologies sales or service office.
617 Configuration error; Could not start license system.
The Software License System has failed to operate. The software option
cannot be enabled. Report this error to the nearest Agilent Technologies
sales or service office.
Configuration error; Data Generator Memory configuration does
not match installed board.
The memory configuration for an option board does not match the known
memory limits of the board. If this error has occurred as the result of a
customer-installed option, uninstall all options and then reinstall the
correct options. Report this error to the nearest Agilent Technologies sales
or service office.
Configuration error; Illegal combination of installed option
boards.
An option board or board combination that is not supported by the
installed firmware has been specified or detected. Be sure that you have
the proper instrument and firmware to support the option board, and that
you have specified a supported combination of options.
Configuration error; Incorrect demodulator board revision.
An invalid Demodulator board is installed.
Configuration error; Incorrect downconverter board revision.
An invalid Downconverter board is installed.