52 ErrorMessages
-199 to -100: Command Errors ESG Family Signal Generators
-128 Numeric data not allowed
A legal numeric data element was received, but the device does not accept
one in this position for the header.
-124 Too many digits
The mantissa of a decimal-numeric data element contained more than 255
digits excluding leading zeros (see IEEE 488.2, 7.7.2.4.1).
-123 Exponent too large
The magnitude of an exponent was greater than 32000 (see IEEE 488.2,
7.7.2.4.1).
-121 Invalid character in number
An invalid character for the data type being parsed was encountered. For
example, an alpha in a decimal numeric or a “9” in octal data.
-120 Numeric data error
This error, as well as errors -121 through -129, are generated when
parsing a data element which appears to be numeric, including
non-decimal numeric types. This particular error message is used if the
device cannot detect a more specific error.
-114 Header suffix out of range
The value of a header suffix attached to a program mnemonic makes the
header invalid.
-113 Undefined header
The header is syntactically correct, but it is undefined for this specific
device. For example, *XYZ is not defined for any device.
-112 Program mnemonic too long
The header contains more than twelve characters (see IEEE 488.2,
7.6.1.4.1).
-111 Header separator error
A character that is not a legal header separator was encountered while