Agilent Technologies E4400-90333 Portable Generator User Manual


 
Error Messages 29
ESG Family Signal Generators -299 to -200: Execution Errors
Too much data; Only 16 distinct values are allowed for I and Q
values.
Option UN8 cannot use an I/Q map that has more than 16 unique values
defined for I and Q (distinct values are listed in the I/Q table editor).
You can use global search and replace to replace all occurrences of a given
value (in both I and Q) with a different value.
Too Much Data; Only 16 distinct values are allowed for I and Q
values.
Although up to 256 I and Q pairs can be configured, only 16 unique I and
16 unique Q values are allowed. If there are more than 16 I and Q pairs,
some of the pairs must use I and Q values that are also used by other
pairs. For example, two I and two Q values can still define four unique I
and Q pairs.
Too Much Data; Only 16 distinct values are allowed for I and Q
values.
The current hardware only allows a maximum of 16 unique values for I
and Q. Try a different modulation.
Too Much Data; PRAM download exceeds the size of PRAM memory.
Download aborted.
You attempted to download a data block to PRAM memory that was larger
than PRAM memory. Either use a smaller pattern, or get more memory by
ordering the appropriate hardware option. Option UN8 has 1 Mbyte of
PRAM, which can store 1 Mbit of data bits. Option UN8 with UN9
contains 8 Mbytes of PRAM, which can store 8 Mbits of data bits.
Too much data; User filter has too many symbols.
Option UN8 cannot use a filter that has more than 32 symbols. You may
have specified an incorrect over-sample ratio in the filter table editor.
-222 Data out of range
A legal program data element was parsed but could not be executed
because the interpreted value was outside the legal range defined by the
device (see IEEE 488.2 11.5.1.1.5).
Data out of range; Channel type not allowed for current
link/config.
The channel type you are trying to add is not supported by the current
cdma2000 link and/or radio configuration.